Title: Reproducibility of retinal nerve fiber layer thickness measurements with specteral domain Cirrus OCT in glaucomatous eyes
Author(s): Reza soltani moghadam,MD .Yousef Alizadeh,MD . Mohamad Javad Mohamadi,MD. Mohamadreza Panjtan Panah, MD. Mohamadreza Absari Haghighi,Ehsan Kazemnejad
Presentation Type: Oral
Subject: Glaucoma
Others:
Presenting Author:
Name: Reza Soltani moghaddam
Affiliation :(optional) Guilan university of medical sciences, Eye research center
E mail: Reza_sm76@yahoo.com
Phone: 02614426908
Mobile: 09123602196
Purpose:

To assess the reproducibility of spectral domain Cirrus high-definition optical coherence tomography (HD-OCT; Carl Zeiss Meditec, Dublin, CA, USA) for analysis of peripapillary retinal nerve fiber layer (RNFL) thickness in glaucomatous eyes

Methods:

Fourty five eyes (one eye from each glaucomatos patient) were imaged with Cirrus HD-OCT. Each eye was imaged three times by two separate operators. Intraclass correlation coefficient (ICC), coefficient of variation (COV), and test–retest variability were evaluated for both intraobserver and interobsever measurements

Results:

In intraobserver measurements, the average thickness ICC was 0.983. CV and test–retest variability were 2.3% and 4.4 µ respectively. In quadrants ICC ranged from 0.886 to 0.956, the lowest associated with nasal quadrant and CV ranged from 3.6 to 7.7%. In interobsever measurements, the average thickness ICC was 0.979. CV and test–retest variability were 2.4% and 4.5 µ respectively. In quadrants ICC ranged from 0.886 to 0.957, the lowest associated with nasal quadrant and CV ranged from 3.8 to 8.6%

Conclusion:

The reproducibility of Cirrus OCT for RNFL thickness is sufficiently good to be useful clinically as a measure of glaucoma progression

Attachment:





Skip Navigation Links
        صفحه اصلی
        ساختار کنگره
        تاریخ های مهم
        مکان برگزاری
        ثبت نام شدگان
        اطلاعات هتل - رزرو
        ارسال مقالات
        برنامه کنگره
        برنامه های جانبی
        نمایشگاه
        تماس با ما
        آرشیو کنگره سالهای گذشته
        جستجوی سخنران
 
Last News

  - بـیــست و سومین کنــگــره سـالیـانه انـجـمـن چـشـم پـزشـکی ایـــران